RephiLe BioscienceDirect-Pure® EDi 5 Water System
the product water is above 5 MΩ·cm at 250C, which meets or exceeds Type II water quality as defined by ASTM®, CAP, CLSI and ISO 3696/BS 3997and also complies with the Purified Water requirements from the European and U.S. Pharmacopoeia.
Features
I. The system incorporates the best in class Siemens EDI module to deliver high quality Type II water consistently at relatively lower cost.
II. Automatic water quality control programs ensure EDI product water meeting or exceeding Type II water requirements.
III. The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane life monitoring as well as many other programs for convenient operation.
IV. The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users to set up additional cleaning procedures.
V. Users can set water dispensing time for easy water collection.
VI. The system is equipped with a 30-liter PE tank (upgradable to a 60-liter) and a liquid level sensor and a tank vent filter with CO2 scavenger to prevent EDI water from carbon dioxide contamination.
VII. RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.
Main Applications
Type II Pure water can be used in many areas. Some of the typical applications are.
• Preparation of bio-reagents
• Preparation of culture media
• Preparation of solution for chemical analysis such as HPLC, ICP etc.
• For clinical analyzers
•Feed water for ultrapure water systems
• Medical device and equipment rinsing
• For serum and blood fractions
• For ophthalmics
Specifications
Feed water quality |
|
Tap water* |
Municipal water, TDS <1000 ppm (2000 μS/cm) |
Water temperature |
5~45 ℃ |
Water pressure |
15~60 psi (1-4 bar) |
Product Water Flow Rate |
|
Type II water |
Typically 5 L/h (at 25℃ ) |
Product Water Quality |
|
Resistivity |
>5 MΩ.cm @ 25°C (typically > 10 MΩ.cm @ 25°C) |
TOC |
< 30 ppb |
Particles in Ultrapure Water (>0.2 μm) |
< 1 /mL (with a 0.2μm final filter) |
Microorganisms |
< 1 cfu/ml (with a 0.2μm final filter) |
Pyrogens (endotoxins) |
< 0.001Eu/ml (with a terminal ultrafiltration cartridge) |
Dimensions |
|
Length x Depth x Height |
30 cm x 49 cm x 48 cm |
Net Weight (Operating Weight) |
22~25 kg (25~32 kg) |
*If water hardness is greater than 1000ppm, RephiLe suggests using an additional water softening device to protect the RO membrane in your system.
Ordering Information
Description |
Catalogue No. |
Direct-Pure EDI 5 Pure Water System |
RD0E00500 |
RO membrane, 1/PK |
RAR005001 |
Polypropylene Depth Filter, 10 Inches 1 µm, 12/PK |
LABPA0112 |
Polypropylene Depth Filter, 10 Inches 10 µm, 12/PK |
LABPA1012 |
Activated Carbon cartridge, 10 Inches, 12/Pk |
LABAC1012 |
Softening cartridge, 10 Inches, 12/Pk |
LABAT1012 |
Tank Vent Filter with CO2 Remover |
RATANKVN1 |
30-liter PE tank |
RATANK030 |
60-liter PE tank |
RATANK060 |