RephiLe Bioscience Direct-Pure® adept Ultrapure Lab Water System
The small foot print Direct-Pure adept lab water system is an ideal choice for users who need up to 20 liters of ultrapure water or RO water per day. Quality of ultrapure product water meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards. |
Features
- No external storage tank is required, which eliminate the risk of water degration by contaminants from the container or the tubing.
- The fully automatic controlled and maintained system contains automatic cleaning cycles and self-maintenance functions for the RO membrane, which increases the lifetime and efficiency
of the RO membranes as well as the cartridges.
- Automatic hourly system recirculation in the loop prevents bacterial growth and maintain water purity at all times.
- System status and key quality parameters are clearly displayed on large LCD screen for easy system performance monitoring.
- Preparation of reagent blank solution
- As sample diluents for GC, HPLC, AA, ICP-MS and other analytical techniques
- Preparation of buffer and culture media for mammalian cell culture
- Preparation of molecular biology reagents, etc.
Specifications
Feed water quality |
|
Tap water* |
Municipal water, TDS <1000 ppm (2000 μS/cm) |
Water temperature |
5~45 ℃ |
Water pressure |
15~60 psi (1-6 bar) |
Product Water Flow Rate |
|
RO water |
> 0.5 L/min at 25℃ |
Ultrapure water (Type I) | > 0.5 L/min at 25℃ |
Product Water Quality |
|
Resistivity |
18.2 MΩ·cm at 25℃ |
TOC* |
< 10 ppb |
Particles in Ultrapure Water (>0.2 μm) |
No particles (with a 0.2 μm final filter or RephiBio filter) |
Microorganisms |
< 0.01 cfu/ml (with a 0.2 μm final filter or RephiBio filter) |
Pyrogens (endotoxins) |
< 0.001 Eu/ml (with a RephiBio filter) |
RNase | < 0.5 pg/ml (with a RephiBio filter) |
DNase | < 10 pg/ml (with a RephiBio filter) |
Dimensions |
|
Length x Depth x Height |
30 cm x 58 cm x 48 cm |
Net Weight (Operating Weight) |
22~25 kg (25~32 kg) |
* Product water quality may vary due to local feed water conditions.
* If water hardness is greater than 1000 ppm, RephiLe suggests using an additional water softening device in order to protect the RO membrane in your system.
Ordering Information
Description |
Direct-Pure Water System, adept |
Direct-Pure Water System, adept UV |