RephiLe Bioscience Large Direct-Pure EDI Water Systems

The large Direct-Pure® EDI system produces Type II water from tap water directly. Resistivity of product water is above 5 MΩ·cm (25oC) ,which meets or exceeds Type Ⅱ water quality standard defined by ASTM®, CAP, CLSI and ISO® 3696/BS 3997, the European and U.S. Pharmacopoeia. The System is designed for users who need 400 to 1800 liters of Type II water per day.

Features

The system incorporates the best in class Siemens EDI module to deliver high quality Type II water consistently at relatively lower cost.

Automatic water quality control programs ensure EDI product water meeting or exceeding Type II water requirements.

The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane life monitoring as well as many other programs for convenient operation.

The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users to set up additional cleaning procedures.

Users can set water dispensing time for easy water collection.

RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements. 

 

Main Applications

Type II Pure water can be used in many areas. Some of the typical applications are.

Preparation of bio-reagents

Preparation of culture media

Preparation of solution for chemical analysis such as HPLC, ICP etc.

For clinical analyzers

Feed water for ultrapure water systems

Medical device and equipment rinsing

For serum and blood fractions

For ophthalmics

 

Specifications

Large Direct-Pure® Water System, EDI

Feed water quality

Tap water*

Municipal water, TDS <1000 ppm (2000 μS/cm)

Water temperature

545 oC

Water pressure

3090 psi (2-6 bar)

Product Water Flow Rate              Typically 30, 60 or 125 L/h (at 25 oC)

Product Water (Type II ) Quality

Resistivity

>5 MΩ.cm @ 25 oC

(typically > 10 MΩ.cm @ 25°C)

TOC

< 30 ppb

Particles in Ultrapure Water (>0.2 μm)

< 1 /mL (with a 0.2μm final filter)

Microorganisms

< 1 cfu/ml (with a 0.2μm final filter)

Pyrogens (endotoxins)

< 0.001Eu/ml 

(with a terminal ultrafiltration cartridge)

Dimensions

Length x Depth x Height

37 cm x 61 cm x 110cm

* If water hardness is greater than 1000 ppm, RephiLe suggests using an additional water softening device

  in order to protect the RO membrane in your system.

 

Ordering Information

Description

Catalogue No.

Direct-Pure Water System, EDI 30

RD0E03000

Direct-Pure Water System, EDI 60

RD0E06000

Direct-Pure Water System, EDI 125

RD0E01H00