RephiLe Bioscience Direct-Pure® UP10 Ultra Pure & 2-Pass RO Water System

Direct-Pure® UP10 Ultra Pure and 2-Pass RO water system is an integrated system that can produce ultrapure and 2-Pass reverse osmosis (RO) water from tap water directly. Quality of ultrapure water produced meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.

Features

━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━

Electrical and water compartments are completely segregated, thus safer to operate.

The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane

   and cartridge life monitoring as well as many other programs for convenient operation.

The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users

    to set up additional cleaning procedures.

Two sets of dual-column purification cartridges work sequentially to ensure thorough removal of ions and organics in water.

The system recirculates automatically every hour to prevent microbial from growing

Resistivity is measured by a high-precision resistivity meter.

Users can set water dispensing time for easy water collection.

The system is equipped with a 30-liter PE tank(upgradable to a 60-liter) and a liquid level sensor,.

A 0.2 μm final filter or ultrafiltration filter (optional) removes final contaminants.

RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.

 

Main Applications

━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━

Pure water can be used in many areas. Here are some typical applications:

Ultrapure Water

RO Pure Water

Applications

Applications

HPLC mobile phase preparation

Preparation of reagent blank solution

As sample diluents for GC, HPLC, AA, ICP-MS and

 other analytical techniques

Preparation of buffer and culture media for mammalian cell culture

Preparation of molecular biology reagents, etc.

Glassware cleaning

Washing machine for glassware

Water bath water

Autoclave

.

Specifications

━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━

 

Feed water quality

Tap water*

Municipal water, TDS <1000 ppm (2000 μS/cm)

Operating temperature

5~45

Feed water pressure

15~60 psi (1.0~4.0 bar)

Product Water Flow Rate                

Ultrapure water (Type I)

Typically 1.5 L/min (at 25 )

RO water

Typically 10 L/h (at 25)

Product Water Quality

Resistivity

   >18.2 MΩ.cm at 25°C

TOC

<10ppb<5 ppb with a dual wavelength UV lamp

Particles in Ultrapure Water (>0.2 μm)

< 1 /mL  (with a 0.2µm final filter)

Microorganisms

< 1 cfu/ml  (with a 0.2µm final filter)

Pyrogens (endotoxins)

< 0.001Eu/ml (with a terminal ultrafiltration cartridge)

Dimensions

Length x Depth x Height

30 cm x 49 cm x 48 cm

Net Weight (Operating Weight)

2225 kg (2532 kg)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

* If water hardness is greater than 1000 ppm, RephiLe suggests using an additional water softening device in order to protect the RO membrane in your system.

** Product water quality may vary due to local feedwater conditions.

 

Ordering Information

━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━  

Description

Catalogue No

Direct-Pure System

 

Direct-Pure UP 10 Ultrapure and 2-Pass RO Water System

RD0P21000

Consumables

Catalogue No.

RephiDuo H Pack cartridge

RR300Q201

RephiDuo U Pack cartridge

RR600Q201

RephiDuo U Pack cartridge with a 0.2 μm PES Capsule Filter

RRQ2F6SKT

RO membrane for Direct-Pure UP 10 System, 1/PK

RAR010001

RO membrane for Direct-Pure UP 2-Pass 10 System, 1/PK

RAR020001

Polypropylene Depth Filter, 10 Inches 1 µm, 12/Pk

LABPA0112

Polypropylene Depth Filter, 10 Inches 10 µm, 12/Pk

LABPA1012

Activated Carbon cartridge, 10 Inches, 12/Pk

LABAC1012

Softening carridge, 10 Inches, 12/Pk

LABAT1012

0.2 μm PES Final Capsule Filter

RAFFC0250

30-liter PE tank

RATANK030

60-liter PE tank

RATANK060

 

 

 

 

 

 

 

 

 

 

 

 

 

Back to Ultra-Pure water systems main page

 

 

Menu