RephiLe Bioscience Direct-Pure® UP10 Ultra Pure & 2-Pass RO Water System
Direct-Pure® UP10 Ultra Pure and 2-Pass RO water system is an integrated system that can produce ultrapure and 2-Pass reverse osmosis (RO) water from tap water directly. Quality of ultrapure water produced meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards. |
Features
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
• Electrical and water compartments are completely segregated, thus safer to operate.
• The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane
and cartridge life monitoring as well as many other programs for convenient operation.
• The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users
to set up additional cleaning procedures.
• Two sets of dual-column purification cartridges work sequentially to ensure thorough removal of ions and organics in water.
• The system recirculates automatically every hour to prevent microbial from growing
• Resistivity is measured by a high-precision resistivity meter.
• Users can set water dispensing time for easy water collection.
• The system is equipped with a 30-liter PE tank(upgradable to a 60-liter) and a liquid level sensor,.
• A 0.2 μm final filter or ultrafiltration filter (optional) removes final contaminants.
• RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.
Main Applications
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Pure water can be used in many areas. Here are some typical applications:
Ultrapure Water |
RO Pure Water |
Applications |
Applications |
• HPLC mobile phase preparation • Preparation of reagent blank solution • As sample diluents for GC, HPLC, AA, ICP-MS and other analytical techniques • Preparation of buffer and culture media for mammalian cell culture • Preparation of molecular biology reagents, etc. |
• Glassware cleaning • Washing machine for glassware • Water bath water • Autoclave . |
Specifications
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Feed water quality |
|
Tap water* |
Municipal water, TDS <1000 ppm (2000 μS/cm) |
Operating temperature |
5~ |
Feed water pressure |
15~60 psi (1.0~4.0 bar) |
Product Water Flow Rate |
|
Ultrapure water (Type I) |
Typically |
RO water |
Typically |
Product Water Quality |
|
Resistivity |
>18.2 MΩ.cm at |
TOC |
<10ppb(<5 ppb with a dual wavelength UV lamp) |
Particles in Ultrapure Water (>0.2 μm) |
< 1 /mL (with a 0.2µm final filter) |
Microorganisms |
< 1 cfu/ml (with a 0.2µm final filter) |
Pyrogens (endotoxins) |
< 0.001Eu/ml (with a terminal ultrafiltration cartridge) |
Dimensions |
|
Length x Depth x Height |
|
Net Weight (Operating Weight) |
22~ |
* If water hardness is greater than 1000 ppm, RephiLe suggests using an additional water softening device in order to protect the RO membrane in your system.
** Product water quality may vary due to local feedwater conditions.
Ordering Information
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Description |
Catalogue No |
Direct-Pure System |
|
Direct-Pure UP 10 Ultrapure and 2-Pass RO Water System |
RD0P21000 |
Consumables |
Catalogue No. |
RephiDuo H Pack cartridge |
RR300Q201 |
RephiDuo U Pack cartridge |
RR600Q201 |
RephiDuo U Pack cartridge with a 0.2 μm PES Capsule Filter |
RRQ |
RO membrane for Direct-Pure UP 10 System, 1/PK |
RAR010001 |
RO membrane for Direct-Pure UP 2-Pass 10 System, 1/PK |
RAR020001 |
Polypropylene Depth Filter, 10 Inches 1 µm, 12/Pk |
LABPA0112 |
Polypropylene Depth Filter, 10 Inches 10 µm, 12/Pk |
LABPA1012 |
Activated Carbon cartridge, 10 Inches, 12/Pk |
LABAC1012 |
Softening carridge, 10 Inches, 12/Pk |
LABAT1012 |
0.2 μm PES Final Capsule Filter |
RAFFC0250 |
30-liter PE tank |
RATANK030 |
60-liter PE tank |
RATANK060 |
Back to Ultra-Pure water systems main page