RephiLe Bioscience Direct-Pure® EDI UV 5 Pure Water System

Direct-Pure® EDI UV 5 lab water system produces Type II water from tap water directly. Resistivity of

the product water is above 5 MΩ·cm at 250C, which meets or exceeds Type II water quality as defined by ASTM®, CAP, CLSI and ISO 3696/BS 3997and also complies with the Purified Water requirements from the European and U.S. Pharmacopoeia.

Features

I.    The system incorporates the best in class Siemens EDI module to deliver high quality Type II water consistently at 
        relatively lower cost.
II.    Automatic water quality control programs ensure EDI product water meeting or exceeding Type II water requirements.
III.    The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane life 
        monitoring as well as many other programs for convenient operation.
IV.   The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users to set
       up additional cleaning procedures.
V.   Users can set water dispensing time for easy water collection.
VI.   The system is equipped with a 30-liter PE tank (upgradable to a 60-liter) and a liquid level sensor and a tank vent filter with
       CO2 scavenger to prevent EDI water from carbon dioxide contamination.
VII.  RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.

Main Applications

Type II Pure water can be used in many areas. Some of the typical applications are.

Preparation of bio-reagents

Preparation of culture media

Preparation of solution for chemical analysis such as HPLC, ICP etc.

For clinical analyzers

Feed water for ultrapure water systems

Medical device and equipment rinsing

For serum and blood fractions

For ophthalmics

 

Specifications

Feed water quality

Tap water*

Municipal water, TDS <1000 ppm (2000 μS/cm)

Water temperature

545

Water pressure

1560 psi (1-4 bar)

Product Water Flow Rate

Type II water

Typically 5 L/h (at 25 )

Product Water Quality

Resistivity

>5 MΩ.cm @ 25°C (typically > 10 MΩ.cm @ 25°C)

TOC

< 30 ppb

Particles in Ultrapure Water (>0.2 μm)

< 1 /mL (with a 0.2μm final filter)

Microorganisms

< 1 cfu/ml (with a 0.2μm final filter)

Pyrogens (endotoxins)

< 0.001Eu/ml (with a terminal ultrafiltration cartridge)

Dimensions

Length x Depth x Height

30 cm x 49 cm x 48 cm

Net Weight (Operating Weight)

2225 kg (2532 kg)

 

* If water hardness is greater than 1000 ppm, RephiLe suggests using an additional water softening device in order to protect the RO membrane in your system.

 

Ordering Information

 

Description

Catalogue No.

Direct-Pure EDI UV 5 Pure Water System

RD0E005UV

RO membrane, 1/PK

RAR005001

Polypropylene Depth Filter, 10 Inches 1 µm, 12/PK

LABPA0112

Polypropylene Depth Filter, 10 Inches 10 µm, 12/PK

LABPA1012

Activated Carbon cartridge, 10 Inches, 12/Pk

LABAC1012

Softening cartridge, 10 Inches, 12/Pk

LABAT1012

Tank Vent Filter with CO2 Remover

RATANKVN1

30-liter PE tank

RATANK030

60-liter PE tank

RATANK060

 

Back to EDI & RO water systems main page

 

Menu